Actosed® Forte Ready provides high-level disinfection for endoscopes and medical instruments. Actosed® Forte Ready solution can be used for up to 15 days. The minimum effective concentration should be checked with the test strips.
Actosed® Forte Ready can be used manually or with an endoscope washing machine for high-level disinfection of flexible and rigid endoscopes, surgical instruments made of stainless steel, including Minimally Invasive Surgical (MIS) instruments, anesthesia materials and non-heat-resistant medical instruments.
Microbiological Activity | Test Method | Concentration | Effect Time |
---|---|---|---|
Bakterisidal | EN 14561 | Ready-to-use | 15 minute |
Fungicidal (C. albicans, A. brasiliensis) | EN 14562 | Ready-to-use | 15 minute |
Tuberculocidal, Mycobactericidal | EN 14348 / EN 14563 | Ready-to-use | 15 minute |
Virucidal (Adeno, Polio, MNV) | EN 14476 | Ready-to-use | 15 minute |
Sporocidal | EN 13704 | Ready-to-use | 1 hours |
Actosed® Forte Ready is suitable for medical instruments and endoscopes made of glass, ceramic, stainless steel, hard rubber as well as synthetic materials such as hard plastic, rubber (including latex and silicone), polyacrylate and polycarbonate. The product is used in the widely-known endoscope brands such as Wolf, Storz, Fujinon, Pentax, Olympus.
Actosed® Forte Ready is ready for use. Before using Actosed® Forte Ready, clean the instrument surfaces and canals with a suitable enzymatic cleaner. (We recommend Actoclean® Perfect). After cleaning, rinse and dry the instruments. Pour
Actosed® Forte Ready solution into the instrument bath. Make sure that all instruments and ducts come into contact with the solution during application. Refer to the microbiological activity table for recommended contact times. Keep the lid of the disinfection bath closed during use. Actosed® Forte Ready solution is stable for up to 15 days. The minimum effective concentration (MEC) of the product used should be checked with the test strips. If turbidity occurs in the appearance of the solution and if the contamination increases, it should be replaced before 15 days. For use with the endoscope washing machine, always follow the device manufacturer’s instructions. For disposal, follow the procedures indicated on the label completely.
After the recommended contact time, rinse the medical instrument with water of at least drinking water quality and dry. During the rinse, use enough water to completely cover the instrument and leave the instrument in the water for at least 1 minute. Repeat this process 2 (two) more times. Follow the manufacturer’s instructions for additional rinsing instructions.
Actosed® Forte Ready is suitable for medical instruments and endoscopes made of glass, ceramic, stainless steel, hard rubber as well as synthetic materials such as hard plastic, rubber (including latex and silicone), polyacrylate and polycarbonate.The product is used in the widely-known endoscope brands such as Wolf, Storz, Fujinon, Pentax, Olympus.
Article-No. | Unit | Pieces in Box |
---|---|---|
01.0303.5 | 5 L | 4 |
14.4164.25 | Test Strip – Pack of 25 pcs | 1 |